Instrument Functions
NRP2
130
User Manual 1173.9140.02 ─ 07
The "Config" tab contains the threshold parameters to define the high, mid and low ref-
erence levels which are used to determine the pulse timing. All values are specified in
percent of the pulse amplitude, either expressed in terms of power (Watt) or voltage
(Volt).
"Low Ref."
The "Low Ref." level defines the start of the rising edge and the end
of the falling edge of the pulse, needed for measurement of the rise /
fall time.
"High Ref."
The "High Ref." level defines the end of the rising edge and the start
of the falling edge of the pulse, needed for measurement of the rise /
fall time.
"Reference"
This reference level is used to define the pulse width, pulse start time
and pulse stop time.
"Ref's related
to"
Selects how the threshold parameters are to be interpreted, either
voltage related or power related. Changing from a power based defi-
nition to a voltage based definition and vice versa affects the analy-
sis, because different points of the waveform are analyzed.
The following table gives the % settings related to power and the dB
figures if the same results are to be obtained as with a typical
10/50/90 definition in terms of voltage. The log values are shown for
comparison only.
Reference Level
Voltage related
(%V)
Power related
(%W)
Log. scale (dB)
Low Ref.
10
1
-20
Ref.
50
25
-6
High Ref.
90
81
-0.9
The voltage related definition represents the normal case for analyz-
ing the pulse parameters.