Interfaces and Connectors
R&S
®
SFE100
24
Getting Started 2112.4122.62 ─ 13
Do not connect or disconnect the network cable until the instrument is switched off.
Otherwise, the network connection cannot be reliably detected.
Electromagnetic interference (EMI) can affect the measurement results. To avoid any
impact, use category 7 cables.
For further information:
●
●
See user manual or the help system for remote control using the Ethernet
Table 5-3: Pin assignment
Pin
Abbreviation
Signal
1
TXD+
Transmit data, positive
2
TXD-
Transmit data, negative
3
RXD+
Receive data, positive
4, 5
R0
Termination, 75R
6
RXD-
Receive data, negative
7, 8
R1
Termination, 75R
5.2.7 DIG I/Q
Requires the extended I/Q input option (R&S
SFE100-K80). Used to receive digital I/Q
signals from another R&S
instrument with the same interface.
For further details see the user manual or the help system.
5.2.7.1
DIG I/Q IN
Digital I/Q input of the R&S
SFE100.
Rear Panel