Remote control commands
R&S
®
RTM3000
640
User Manual 1335.9090.02 ─ 09
BUS<b>:CAN:SAMPlepoint
<SamplePoint>
Sets the position of the sample point within the bit in percent of the nominal bit time.
See also:
Suffix:
<b>
.
1..4
Parameters:
<SamplePoint>
Range:
10 to 90
Increment: 1
*RST:
50
Default unit: %
BUS<b>:CAN:BITRate
<BitRate>
Sets the number of transmitted bits per second.
Suffix:
<b>
.
1..4
Parameters:
<BitRate>
Range:
100 to 5,04E06
Increment: Depends on the bit rate value
*RST:
50E03
Default unit: Bit/s
17.11.5.2
CAN - trigger
To configure the protocol trigger, make sure to set first:
●
BUS
●
SBUS1 | SBUS2
............................................................................................ 642
.............................................................................................. 642
...........................................................................................642
...............................................................................................643
............................................................................................. 643
.............................................................................................. 643
............................................................................................644