Remote control commands
R&S
®
RTM3000
577
User Manual 1335.9090.02 ─ 09
17.9.1.5
Masks
In addition to the commands described below, consider also the following commands:
●
●
●
●
●
MASK:DATA?
Returns the data of the mask. The mask consists of two limit curves.
Return values:
<Data>
List of values according to the format settings - the y-values of
the mask points. The list contains two values for each sample
interval.
Usage:
Query only
MASK:DATA:HEADer?
Returns information on the mask data that is delivered with
.
Table 17-5: Header data
Position
Meaning
Example
1
XStart in s
-9.477E-008 = - 94,77 ns
2
XStop in s
9.477E-008 = 94,77 ns
3
Number of samples
200000
4
Number of values per sample interval. For masks
the value is 2.
2
Return values:
<DataHeader>
Comma-separated value list
Example:
-9.477E-008,9.477E-008,200000,2
Usage:
Query only
17.9.1.6
Logic channels
See
Chapter 17.13.1.3, "Logic channels - waveform data"
Documenting results