Remote control commands
R&S
®
RTM3000
446
User Manual 1335.9090.02 ─ 09
PROBe<m>:SETup:BANDwidth?
Returns the bandwidth of the probe.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Return values:
<Bandwidth>
Range:
10e5 to 20e8
Increment: 10
Default unit: Hz
Usage:
Query only
PROBe<m>:SETup:ATTenuation[:AUTO]?
Returns the attenuation of an automatically detected probe.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Return values:
<ProbeAttenuation> Range:
0.0001 to 10000
Usage:
Query only
PROBe<m>:SETup:GAIN[:AUTO]?
Returns the gain of an automatically detected probe.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Return values:
<ProbeGain>
Range:
0.0001 to 10000
Usage:
Query only
PROBe<m>:SETup:CAPacitance?
Returns the input capacity of the probe.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Waveform setup