Basic Operation
R&S
®
Cable Rider ZPH
83
User Manual 1321.0950.02 ─ 07
To measure the harmonic ratio, set the marker on the signal and a delta marker on
the second harmonic.
9. Press the [MARKER] key.
The R&S Cable Rider ZPH sets a marker on the trace maximum. The trace maxi-
mum corresponds to the signal.
10. Select the "New Marker" softkey.
The R&S Cable Rider ZPH activates a delta marker and places it on the next trace
maximum. This corresponds to the second harmonic.
The harmonic ratio is the vertical distance of the marker and the delta marker. The
R&S Cable Rider ZPH displays this value in the
Measurement Result View
.
4.8
Using a Power Sensor
R&S Cable Rider ZPH option
R&S ZPH-K9 (order number: 1321.0409.02) option is required to operate the R&S
Cable Rider ZPH in the power sensor mode.
For highly accurate power measurements, you can connect one of the power sensors
that are supported by the R&S Cable Rider ZPH.
For a list of R&S Cable Rider ZPH supported power sensor, refer to
Chapter 9.1,
"Using a Power Sensor"
,
on page 253.
You can connect the list of supported power sensors to the USB port of the R&S Cable
Rider ZPH. This connector allows you to control the power sensor and supplies it with
power. For more information, see
"USB type A connector"
on page 28.
Using a Power Sensor