Remote Commands
R&S
®
Cable Rider ZPH
460
User Manual 1321.0950.02 ─ 07
[SENSe:]FREQuency:CENTer
<CenterFrequency>
This command defines the center frequency of the R&S Cable Rider ZPH.
In spectrum analyzer mode, the command also defines the measuring frequency for
time domain measurements (span
=
0).
Parameters:
<CenterFrequency> Range:
Depends on the operating mode and is specified in
the data sheet.
*RST:
fmax /2 with fmax = maximum frequency
Default unit: MHz
Example:
FREQ:CENT 100MHz
Defines a center frequency of 100 MHz.
[SENSe:]FREQuency:CENTer:STEP
<Stepsize>
This command defines the center frequency step size.
Parameters:
<Stepsize>
Range:
1 Hz to fmax
*RST:
- (AUTO 0.1*SPAN is switched on)
Default unit: MHz
Example:
FREQ:CENT:STEP 120MHz
Defines a CF step size of 120 MHz.
14.6.4.2
Defining the Vertical Axis
The following chapter describes command necessary to define the vertical scale such
as the x-axis deviation per division and settings affecting the amplitude of the measure-
ments.
List of commands
CALCulate<1...2>:MARKer<1...6>:FUNCtion:LEVel:ONCE
.................................................460
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:PDIVision
.......................................... 461
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:RLEVel
..............................................461
INPut:ATTenuation
......................................................................................................... 462
INPut:ATTenuation:AUTO
............................................................................................... 462
INPut:ATTenuation:MODE
...............................................................................................462
INPut:GAIN:STATe
......................................................................................................... 462
INPut:IMPedance
...........................................................................................................463
INPut:IMPedance:PAD
....................................................................................................463
CALCulate<1...2>:MARKer<1...6>:FUNCtion:LEVel:ONCE
This command adjusts the reference level to the measured signal power.
This automatic routine makes sure that the signal power level does not overload the
R&S Cable Rider ZPH or limits the dynamic range by too small an S/N ratio.
Remote Control - Commands