The Measure of Excellence
e-Dial™ P/N: 6317-00001
8
Rochester Gauges, LLC e-Dial™ sensor device is intrinsically safe apparatus intended to
be powered through a designated e-Dial™ Barrier with matching intrinsic safety entity
parameters. e-Dial™ and e-Dial™ Barrier are certified as complete intrinsic safety system
(Reference Intrinsic Safety Control Drawing SD-581) and the maximum supply voltage
to the barrier should not exceed Um=32VDC. e-Dial™ Barrier device is intended to be
installed in safe (non-hazardous) location and e-Dial™ sensor device is intended for
use in Class I, Division 1, Groups C and D, T4 or Zone 0, Group IIB, T4 Hazardous Locations
under the certification schemes and ratings noted below:
IECEx (Global Certification):
e-Dial™ Barrier
[Ex ia Ga] IIB
-40°C ≤ T
AMB
≤ +60°C
e-Dial™
Ex ia IIB T4 Ga
-40°C ≤ T
AMB
≤ +80°C
IECEx ETL 20.0019X
ATEX (EU Certification):
e-Dial™ Barrier
II (1) G [Ex ia Ga] IIB
-40°C ≤ T
AMB
≤ +60°C
e-Dial™
II 1 G Ex ia IIB T4 Ga
-40°C ≤ T
AMB
≤ +80°C
ITS20ATEX205570X
0359
0359
North America (USA & Canada):
e-Dial™ Barrier
To be installed in safe areas and provides
intrinsic safety outputs to:
[Ex ia] Class 1, Division 1, Groups C & D
USA: [AEx ia Ga] IIB
Canada: [Ex ia Ga] IIB
-40°C ≤ T
AMB
≤ +60°C
e-Dial™
Class 1, Division 1, Groups C & D, T4
USA: Class 1, Zone 0, AEx ia IIB T4 Ga
Canada: Ex ia IIB T4 Ga
-40°C ≤ T
AMB
≤ +80°C
ETL20CA104269205X
Содержание 6317-00001
Страница 1: ...User Manual...
Страница 2: ...The Measure of Excellence e Dial P N 6317 00001 ii...