33
Technical data
Mains connection:
Standard:
230V ± 10% 50-60 Hz fuse MT 4T
also available in: 115V ± 10% 50-60 Hz fuse MT 4T
24V ± 10% 50-60 Hz fuse MT 4A
The instrument is nonvolatile
Power consumption
of the analyser:
30 VA
Protective system:
IP
54
Weight:
7
kg
Dimensions:
WxHxD = 300x400x180 mm
Indicators:
Type / measuring range:
CA 3
2.7 -53.4 mg/l CA 9 8 - 160 mg/l
CA 30
27 - 534 mg/l
Consumption per analysis:
Min. quantity per analysis: 0.02ml
Content of indicator bottle:
500 ml
Measured value display:
3-figure,
programmable units: °dH. °E, °F, mg/l Ca CO
3,
gpg and mmol/ltr
Relay outputs:
Number: 3
Electrical data:
floating change-over contact, loadable with max. 250 V, 4A
Programmable functions (max.3):
Pulse signal, Permanent signal 1, Permanent signal 2,
Analysis runs, Signal relay
Analogue outputs:
2
Output 1:
Status display: Analysis active, Soft water, Hard water,
Refill indicator and Measurement fault
0(4)-20
mA
Output 2:
Water hardness, scalable initial and final values
0(4)-20
mA
Inputs:
Number: 2
Electrical data:
Load of the switching contacts with max. 9 V, 8 mA
Programmable functions (max.2):
Start analysis, Stop analysis, Reset relay/horn,
Water
meter
Measurement cycle:
time-dependent (3-9999 min.), volume-dependent,
external or manual
Flush times:
programmable (10-999 sec.)
Water sample:
Measuring chamber content:
18 ml
Sample temperature:
5 - 45 °C
Condition:
clear, no gas bubbles
Chemical requirements:
pH 4 - 8, iron < 3 ppm, copper < 0.2 ppm
Operating pressure:
0.2 -6 bar
Drainage:
pressureless (open funnel)
Water connections:
Water inflow:
Pressure hose, inner diameter 6-7 mm = ¼"
Water outflow:
Hose, inner diameter 6-7 mm = ¼" (open funnel)
range
measuring
of
End
value
measured
x
ml
approx
315
.
0
.