Alarm Activation and Functions
6-1. Gas alarm activation
A gas alarm is activated when detected gas concentration reaches the preset alarm setpoint.
NOTE
•
The alarm setpoint (first alarm and second alarm) is factory-set. The setting values can be changed
in the maintenance mode (P.30).
•
Although the alarm delay time (standard: 2 seconds) prevents false alarm activation, the delay can
be turned off in the maintenance mode (P.30).
Display operation
<Concentration Display>
In case of over the detection range (Over Scale), "∩∩∩∩" is displayed on the LCD.
<During Power-on>
The LCD is continuously displayed.
<Alarm Display (ALM1: Yellow Backlight), (ALM2: Red Backlight)>
The alarms have two levels. The yellow or red backlight lights up when the respective alarm setpoint
value is reached to or exceeded. (Default: Backlight is off)
Contact activation
When gas concentration is at or above the alarm setpoint value, the contact is activated. After the gas
concentration falls below the alarm setpoint value, the contact activation resets automatically.
Measuring ranges of 0-2000ppm / 0-5000ppm / 0-10000ppm have a 50 ppm hysteresis.
alarm setpoint value 1000ppm
950ppm
Hysteresis
50ppm
Contact activation
Содержание RI-600
Страница 35: ...35 2 03 LCD Perform the LCD display test 2 04 Not used To 2 00 ...
Страница 43: ...43 2 4b P A Not used 2 4C ACAL Not used 2 4d BZZR Set the buzzer to ON OFF ...
Страница 44: ...44 2 4E AL R Set the gas alarm state contact to ON OFF 2 4F MA O Set the external output signal to ON OFF ...
Страница 45: ...45 2 4H LCD Set the LCD backlight to ON OFF No backlight by standard setting ...
Страница 60: ......
Страница 61: ......