51
8
Connect the cable plug into
the sensor board.
9
Secure the sensor cover of
the new cable’s remote
sensor housing.
10
Route the new remote
cable’s wiring through the
hole in the housing to the
right of the sensor terminal
strip and install a zip tie in
the case and around the
cable as shown in the
image to the right, but do
not tighten it yet.
11
Using the terminal
pushbuttons, connect the
cable’s wires to the sensor
terminal strip. See the table
below for the wire colors.
Type A wire colors are
shown in the figure to the
right. Note that the Type B
cable has two pairs of wires
extending from the cable.
Tighten the zip tie to secure
the cable in place.
Terminal 1
Terminal 2
Terminal 3
Terminal 4
Type A
Red
Green
White
Black
Type B
White wire of
white/yellow pair
Yellow wire of
white/yellow pair
White wire of
white/blue pair
Blue wire of
white/blue pair
12
Secure the top half of the
main unit’s housing to the
bottom half.