How to Install
4-1. Precautions for installation points
When installing the monitor, observe the following precautions. Ignoring these precautions may damage
the monitor, resulting in inaccurate gas detection.
<Do not install the monitor in a place with vibrations or
shocks.>
The monitor consists of sensitive electronic parts and must be installed in a
stable place without vibrations or shocks, etc. and where it cannot fall.
<Do not install the monitor in a place exposed to water, oil, or
chemicals, etc.>
When selecting installation points, avoid a place where the monitor is exposed
to liquids such as water, oil, or chemicals.
<Do not install the monitor in a place where the temperature
drops below 0ºC or rises over 40ºC.>
The operating temperature of the monitor is 0 to +40ºC. Install the monitor in a
stable place not exceeding the operating temperature range.
<Do not install the monitor in a place exposed to direct
sunlight or sudden changes in the temperature.>
Avoid a place where the monitor is exposed to direct sunlight or radiant heat
(infrared rays emitted from a high-temperature object) or where the unit
temperature changes suddenly. Condensation may form inside the monitor,
and the monitor cannot adjust to sudden changes in the temperature.
<Keep the monitor (and its cables) away from noise source
devices.>
When selecting installation points, avoid a place where high-frequency/high-
voltage devices exist.
CAUTION
•
This is a precision device. Because the monitor may not provide the specified performance in
some places (environments), check the installation environment, then take appropriate actions if
necessary. Because the monitor plays an important role for safety and disaster prevention, it
should be installed appropriately.
•
Do not install this product in any of the following locations:
- Outside environments or areas exposed to direct sunlight
- Areas exposed to water
- Areas exposed to ventilation from an air conditioner, etc.
- Areas exposed to soot, smoke, or steam
- Areas where the temperature drops below 0ºC or rises over 40ºC or the temperature changes
suddenly
- Areas with high humidity like a bathroom
- Areas with bad ventilation such as behind a curtain
Содержание RI-600
Страница 35: ...35 2 03 LCD Perform the LCD display test 2 04 Not used To 2 00 ...
Страница 43: ...43 2 4b P A Not used 2 4C ACAL Not used 2 4d BZZR Set the buzzer to ON OFF ...
Страница 44: ...44 2 4E AL R Set the gas alarm state contact to ON OFF 2 4F MA O Set the external output signal to ON OFF ...
Страница 45: ...45 2 4H LCD Set the LCD backlight to ON OFF No backlight by standard setting ...
Страница 60: ......
Страница 61: ......