21
5-4. Power-on
- Before turning on the power switch, check that the monitor is installed properly.
- Slide the power switch up to power on the monitor.
- After the monitor completes the start-up, it enters the detection mode.
<Start-up Procedures (approximately 25 seconds for system check of the monitor and alarm
deactivation)>
Power On
▼
▼
▼
Initial clear
(approx. 25 seconds)
▼
▼
▼
Detection mode
During initial clear
External output
4 - 20 mA: About 2.5 mA fixed
0 - 1 V: About 0 V fixed
Power
switch
ON
OFF
CAUTION
•
Do not turn off the monitor during the initial clear. If it is turned off during the initial clear and
turned on again, abnormal operation may occur.
•
After the initial clear, perform a calibration.
•
In a normal air release environment, CO2 concentration is hundreds of ppm. Since the reading is
not zero, even in the normal state, attention is paid.
•
If there are many people and ventilation is insufficient in the area, CO
2
concentration may exceed
1000ppm (0.1 vol%).
CO
2