EAGLE 2 Operator’s Manual
Data Logging • 53
Instruments, Inc.
The data logging capacity depends on how often the EAGLE 2 stores data, how many
channels are active, and how often the EAGLE 2 is turned on and off. The table below
illustrates how much data logging time is available for the various interval times. It
assumes that the unit is setup with four sensors, is only turned on once, and there are no
alarm occurrences. See “Updating the Data Log Interval Setting” on page 111 for
instructions on setting the data logging interval time.
For a complete description of the Data Logger Management Program and procedures for
downloading data to a computer, see the Eagle 2 Data Logger Management Program
Operator’s Manual.
Table 8: Data Logging Capacity, 4-gas EAGLE 2
Interval
Time
Data Logging Time
5 seconds
239 hours (10 days)
10 seconds
479 hours (20 days)
20 seconds
959 hours (40 days)
30 seconds
1439 hours (60 days)
1 minute
2879 hours (120 days)
3 minutes
8639 hours (360 days)
5 minutes
14,399 hours (600 days)
10 minutes
28,798 hours (2,000 days)