262 • Alarms
EAGLE 2 Operator’s Manual
4. The “S” will then disappear and the following screen will appear for a few seconds.
5. If the unit prompts you for the password, enter it by using the AIR
▲
YES or RANGE
▼
SHIFT buttons to select each password number and then pressing and releasing the
POWER ENTER RESET button to enter it and move on to the next number until all
of the numbers are entered. The main menu displays. It displays six menu items at a
time.
6. Use the AIR
▲
YES or RANGE
▼
SHIFT button to move the cursor to the
ALARM
POINTS
menu item.
7. Press and release POWER ENTER RESET. The Change Alarm Point Settings Screen
appears and all detection channels are displayed.
8. Move the cursor next to the oxygen channel. Press and release POWER ENTER
RESET.
9. A mode select screen will appear and prompt you to choose between Normal Mode
and Inert Mode. Use the AIR
▲
YES or RANGE
▼
SHIFT button to display Inert
Mode and press and release POWER ENTER RESET.
GAS MONITOR
EAGLE 2
>SET DATE & TIME
SET DATE FORMAT
SET BATTERY TYPE
CONFIGURE CHANNELS
CONFIGURE GASES
CATALYTIC UNITS
CHANGE ALARM
POINT SETTINGS
> 1: CH4 2: OXY
3: H2S 4: CO
END
MODE SELECT
UP/DOWN THEN ENTER
NORMAL MODE