43
Preparations
6. Use the
keys to
select [Type] and press
the MENU/ENTER key.
The type screen appears.
7. Use
the
keys to select the type (AC OUT, DC OUT) and press
the MENU/ENTER key.
AC OUT:
The OUTPUT connectors provide AC signal.
DC OUT:
The OUTPUT connectors provide DC signal.
Select the output type that is appropriate for the equipment to be
connected.
Note
It is not possible to select different AC OUT or DC
OUT settings for individual channels.
8. Press the START/STOP key to return to the measurement screen.
9. Press the OUTPUT CAL
key.
The display switches to
calibration indication.
Verify that the measure-
ment value reading is the
same as the maximum
value for the level range
in each axis (X, Y, Z).
During calibration, the OUTPUT connectors supply the following
signal.
AC : 31.5 Hz, 1 Vrms
(OUTPUT setting is “AC OUT” )
DC: 2.5 V
(OUTPUT setting is “DC OUT” )
Use this signal to calibrate the external equipment such as a level
recorder or analyzer.
10. Press the Output Cal key again to cancel the calibration mode.
Содержание VM-55
Страница 2: ......
Страница 14: ...2 Outline Block diagram...
Страница 45: ...33 Preparations System screen Clock settings screen...
Страница 47: ...35 Preparations System screen Backlight LCD settings screen...
Страница 49: ...37 Preparations System screen Execution confirmation screen...
Страница 51: ...39 Preparations Display I O screen Comparator screen...
Страница 141: ...129 Serial Interface Follow the wizard to complete the installation Screens during installation are as follows...
Страница 142: ...130 Serial Interface Depending on your environment Windows Security may be displayed Click on Install or Continue...
Страница 203: ......
Страница 204: ...This product is environment friendly It does not include toxic chemicals on our policy No 60508 19 10...