Store Operation
14. Press the START/STOP key. Measurement will start at the preset
start time. The measurement value will be stored automatically at
every interval set for
L
v
store interval and
L
eq
calculation interval.
When it is completed, the address is incremented by one step.
Measurement will stop at the preset stop time.
To stop the process earlier, press the START/STOP key.
Note
Relationship between elapsed measurement time
and number of data
When using Timer Auto mode and 100-ms sampling,
10 data sets are stored per second. Therefore, when
10 seconds of measurement time have elapsed, the
number of stored data is 100 (10 when using 1-second
sampling).
During Timer Auto mode, the pause function cannot
be used.
During Timer Auto mode, the address indication is
based on number of
L
eq
calculation. The number of
L
eq
calculation is not displayed when
L
eq
calculation
interval is OFF.
If the fi nal measurement would be shorter than the
specifi ed measurement time, that measurement is
not performed. In this case, the operation ends with
the preceding measurement.
Содержание VM-55
Страница 2: ......
Страница 14: ...2 Outline Block diagram...
Страница 45: ...33 Preparations System screen Clock settings screen...
Страница 47: ...35 Preparations System screen Backlight LCD settings screen...
Страница 49: ...37 Preparations System screen Execution confirmation screen...
Страница 51: ...39 Preparations Display I O screen Comparator screen...
Страница 141: ...129 Serial Interface Follow the wizard to complete the installation Screens during installation are as follows...
Страница 142: ...130 Serial Interface Depending on your environment Windows Security may be displayed Click on Install or Continue...
Страница 203: ......
Страница 204: ...This product is environment friendly It does not include toxic chemicals on our policy No 60508 19 10...