7. Maintenance
7-5. How to replace parts
- 70 -
Replacement of Pump Unit
For the replacement of pump unit, see "Detaching and Attaching Pump Unit" in "3-3. Names and functions
for each part".
Replacement of flow sensor, fan and catalyst tube
After the flow sensor, fan or catalyst tube is replaced, the operation must be checked by a qualified service
engineer. Since the catalyst tube has an influence over gas sensitivity, calibration needs to be performed
after replacement. Please contact RIKEN KEIKI.
Sensor Unit Replacement
For the replacement of sensor unit, see "Detaching and Attaching Sensor Unit" in "3-3. Names and
functions for each part". Since the sensor unit has an influence over gas sensitivity, calibration needs to be
performed after replacement.
CAUTION
Turn OFF the power when the sensor unit is replaced.