7. Maintenance
7-3. Calibration method
- 66 -
* When the zero adjustment fails
●
○
○
○
0.0 ppm
ZERO NG
MAINTENANCE
↓
Return to 2-1. ZERO
<Span Adjustment "2-2">
This is used to perform the span adjustment.
PW
A1
A2
F
LCD
2-2. SPAN
Press SET key.
●
○
○
○
2-2
SPAN
MAINTENANCE
↓
↓
Gas introduction
Introduce a gas, and then press the
SET key when the reading is stabilized.
* In case of less than 10%FS, this is not
switched to the next menu.
●
○
○
○
2.0 ppm
SPAN GAS
MAINTENANCE
|
|
↓
↓
Span Adjustment
Because the reading will be fixed,
adjust the introduced gas concentration
by pressing the
▲
or
▼
key. After
adjusting the value, press the SET key.
●
○
○
○
8.0 ppm
SPAN VAL
MAINTENANCE
↓
↓
Span Adjustment Completed
●
○
○
○
8.0 ppm
SPAN OK
MAINTENANCE
↓
↓
Adjustment result record
To record the adjustment result, press
the SET key.
(Press the MODE key to cancel this
menu.)
●
○
○
○
8.0 ppm
SPAN Y/N
MAINTENANCE
↓
↓
CAUTION
Do not press the SET key before the
reading is stabilized.