7. Maintenance
7-2. Regular maintenance mode
- 56 -
▲
↓↑
▼
SET 13. ZERO 24F
This is a supplemental
setting screen for the
above zero follower
function. (A setting to
determine whether the
first-24-hour zero
follower will be
performed after the
power is turned on.)
Do not change the
setting when the gas
detector is used in a
normal way.
(ON setting)
●
○
○
○
SET 13
ZERO 24F
MAINTENANCE
|↑
||
▲
||
▼
||
↓|
SET 14. S ASSIST
Set the sensitivity
correction when the
electrochemical type
(ESU) is installed.
Select either ON/OFF,
and then press the SET
key to confirm the
selection.
When ON is selected,
the sensitivity
correction function is
activated.
●
○
○
○
SET 14
S ASSIST
MAINTENANCE
⇔
SET
ON
S ASSIST
MAINTENANCE
|↑
▲
||
▼
||
↓|
▲↓↑▼
OFF
S ASSIST
MAINTENANCE
SET 15. MNT OUT
Set the external output
for the maintenance
mode.
Select either 2.5 mA/4.0
mA/HOLD (previous
value)/4 - 20 mA (linked
to display value), and
then press the SET key
to confirm the selection.
(4 - 20 mA
specification)
●
○
○
○
SET 15
MNT OUT
MAINTENANCE
⇔
SET
2.5 mA
MNT OUT
MAINTENANCE
|↑
||
||
||
▲
||
▼
||
||
||
↓|
▲↓↑▼
4.0mA
MNT OUT
MAINTENANCE
▲↓↑▼
HOLD
MNT OUT
MAINTENANCE
▲↓↑▼
4-20mA
MNT OUT
MAINTENANCE