7. Maintenance
7-2. Regular maintenance mode
- 64 -
<Catalyst Aging Function “2-13”>
PW
A1
A2
F
LCD
2-13.PL AGING
Press SET key.
●
○
○
○
2-13
PL AGING
MAINTENANCE
↓
▲
↓↑
▼
Then press the SET key
again
for three seconds.
●
○
○
○
SET
PL AGING
MAINTENANCE
↓
↓
The pyrolyzer
temperature is adjusted to
700
º
C automatically.
When 700
º
C is reached, it
is kept for two hours.
●
○
○
○
700
º
C
PL AGING
MAINTENANCE
↓
↓
Two hours later, the
pyrolyzer temperature is
adjusted to the setting
temperature in
measurement
automatically.
* For the COS detector
(120
º
C), the display
looks like the figure on
the right.
●
○
○
○
120
º
C
SET OK
MAINTENANCE
|
|
↓
↓
Two hours later, the aging
of catalyst is completed.
●
○
○
○
End
SET OK
MAINTENANCE