background image

7. Maintenance

 

7-2. Regular maintenance mode 

 

- 53 - 

 

<Environmental Setting 2 "2-10"> 

In the environmental setting 2, specify the settings of functions. (* It is recommended that setting changes 
should be recorded in a log.) 
The environmental setting 2 includes setting menus which are usually not used. Be careful not to change 
these settings by mistake. 
 
 PW 

A1 

A2 

LCD 

 

 

2-10. SETTING2 

Press SET key. 
 

 

 

 

 

2-10 

SETTING2 

MAINTENANCE  

 

 

 

 

 

 

 

 

 

 

SET 0. ADDRESS 

 

* This is not used on 
the gas detector

.

 

 

 

 

 

SET 0 

ADDRESS 

MAINTENANCE

 

 

 

 

 

 

 

↓↑

 

 

 

SET 1. DAY TIME 

Set the date/time of the 
main unit. 

 

 

 

 

SET 1 

DAY TIME 

MAINTENANCE

 

 

SET 

Date and time setting 
=> P58 

 

 

 

 

↓↑

 

 

 

SET 2. SUPPRESS

 

Set the suppression 
value. 
Change the value by 
pressing the 

 or 

 

key, and then press the 
SET key to confirm the 
value. 

 

 

 

 

SET 2 

SUPPRESS 

MAINTENANCE

 

 

SET 

0.9ppm 

SUPPRESS 

MAINTENANCE

 

 

 

 

 

↓↑

 

 

 

SET 3. SUP TYPE

 

Set the suppression 
system. 
Select either 
CUT/SLOPE, and then 
press the SET key to 
confirm the selection. 
When CUT (Cut Off) is 
selected, values 
exceeding the 
suppression value are 
directly displayed. 
When SLOPE (Slope) 
is selected, values 
exceeding the 
suppression value are 
gently displayed. 

 

 

 

 

SET 3 

SUP TYPE 

MAINTENANCE

 

 

SET 

CUT 

SUP TYPE 

MAINTENANCE

 

 

 

 

 

|↑

 

||

 

||

 

↓|

 

 

▲↓↑▼

 

 

 

 

 

 

SLOPE 

SUP TYPE 

MAINTENANCE 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание TP-70DGII

Страница 1: ...PT5E 0581 Gas Detector TP 70DGII Operating Manual...

Страница 2: ...perating procedures 25 5 3 How to activate the gas detector 25 5 4 Modes 28 5 5 Detection mode 29 5 6 Alarm test mode 30 5 7 User mode 31 5 8 How to exit 35 6 Operations and Functions 36 6 1 Gas alarm...

Страница 3: ...t an analyzer or densitometer which performs quantitative qualitative analysis measurement for gases Please fully understand the features of the gas detector before using it so that it can be used pro...

Страница 4: ...labeled on the detector in case of comply with CE marking Please confirm the instrument specification before using Please refer Declaration of Conformity that is at the end of this manual if you have...

Страница 5: ...for defects When seeming defects are found in the protective functions such as protective grounding do not start the gas detector External connection Before connecting the gas detector to a detection...

Страница 6: ...etector in a place where combustible explosive gases or vapors are present Operating the gas detector in such an environment will lead to extreme dangers Response to gas alarm Issuance of a gas alarm...

Страница 7: ...hus condensation must be avoided In addition to the installation environment carefully monitor the temperature humidity of the sampling point to prevent condensation inside the tube In particular when...

Страница 8: ...7 3 3 Product Components 3 1 Main unit and standard accessories Main Unit Standard Accessories Operating Manual Protective rubber cap to be removed when using the gas detector Dust filter Silica gel...

Страница 9: ...lly for the gas detector Do not use a similar product It will not function normally Display unit Sensor unit CAUTION Each unit consists of precision parts When a unit is detached be careful not to dro...

Страница 10: ...front cover This is used to open close the front cover 4 External terminal plate Connect cables for external signals 4 20 mA and contacts 5 Sampling gas inlet outlet GAS IN Connect a sample gas sucti...

Страница 11: ...bnormality is detected in the gas detector 10 Gas name display Displays a gas name in chemical formula etc e g Silane SIH4 11 Concentration value display Displays the gas concentration and so on 12 Un...

Страница 12: ...pull it out If the sensor unit is hard to detach insert a finger into the gap at the top right of the sensor unit and pull it out When attaching the sensor unit to the main unit thrust it onto the ma...

Страница 13: ...ump unit cover With the pump unit cover opened the pump located inside can be pulled out Attach the pump in the reverse order After attaching the pump check that the pump unit cover is fixed securely...

Страница 14: ...r units cannot be stored together with other units Please contact RIKEN KEIKI for more information 3 4 Block diagram Tubing Diagram CAUTION The sensor unit must be handled carefully to ensure quality...

Страница 15: ...ON ALARM Closed contact at an alarm state Alarm contact controller Transmitter NT specification Digital data transmission Ethernet and PoE specifications Ethernet and 4 20 mA DC transmission Display P...

Страница 16: ...ibrations or shocks and it cannot drop CAUTION This is a precision device Because the gas detector may not provide the specified performance in some places environments check the environment in the in...

Страница 17: ...the temperature Keep the gas detector and its cables away from noise source devices When selecting installation points avoid a place where high frequency high voltage devices exist Do not install the...

Страница 18: ...d to the cables may be damaged Lightning cannot be prevented Cables installed in a metal conduit or under the ground cannot be completely protected from inductive lightning surge caused by lightning A...

Страница 19: ...l relay at a lower voltage of 100 VAC or below contact amplification At the same time the surge absorbing part SK1 suitable for the specifications must be attached to the external relay In addition th...

Страница 20: ...100BASE T or higher Cable specification Single wire 5 m maximum for a stranded wire Modular plug RJ 45 Number of cores 8 core Wire connection Straight Figure of Terminal Plate 1 2 3 4 5 6 7 8 9 10 24...

Страница 21: ...on 200 VAC system prepare a power cable that supports 200 VAC The power cable is provided only for the case that the power voltage is 100 VAC system requested The provided power cable supports 100 VA...

Страница 22: ...PLC EA specification with Ethernet Upper unit DCS PLC Upper unit PC PLC Power supply 100 240 VAC Power supply 100 240 VAC RJ45 connector Switching hub CAUTION Do not use a switching hub equipped with...

Страница 23: ...ON Be sure to use a switching hub equipped with PoE function with the PoE specification If the PoE function is unavailable the power will not be supplied to the display unit Upper unit PC PLC Power su...

Страница 24: ...pling inlet and outlet GAS IN GAS OUT of the gas detector detected gases may be leaked from its inside thus leading to dangers Be sure that excessive pressure is not applied to the gas detector while...

Страница 25: ...ive rubber cap from GAS IN and GAS OUT If the gas detector is turned on with the rubber caps remaining while the installation applied overload may damage the pump and sensor Never fail to remove the c...

Страница 26: ...ctor enters the detection mode immediately after the start up flow Warm up Time Unpowered time lasts one week or more Warm up the gas detector for 24 hours or more Note that the warm up time can be re...

Страница 27: ...and then turn on the power switch of the display unit After the gas detector completes the start up it enters the detection mode swiftly Warm up the gas detector for a specified period to obtain accur...

Страница 28: ...the sensor memory during the initial clear When the power is turned off to perform replacement or maintenance of parts the gas detector needs to be warmed up for a specified period which is determine...

Страница 29: ...settings First alarm setpoint AL1 Second alarm setpoint AL2 Alarm delay time Zero suppression value Zero follower ON OFF Sensitivity correction ON OFF Flow Rate Indicator 1 3 FLOW Display the current...

Страница 30: ...hown on the figure below when the flow rate does not correspond to the specified flow rate for some reasons it is adjusted automatically Low flow rate Excess flow rate Normal flow rate CAUTION If the...

Страница 31: ...0 ppm TEST OFF MAINTENANCE First Alarm When the reading reaches the first alarm the ALM1 lamp lights up When Contact ON is selected the first alarm contact is activated 5 0 ppm TEST OFF MAINTENANCE Se...

Страница 32: ...MAINTENANCE SET 0 50 L M MAINTENANCE 1 4 ADDRESS This is not used on the gas detector 1 4 ADDRESS MAINTENANCE 1 5 70D VER Display the program version of the main unit 1 5 70D VER MAINTENANCE SET 01234...

Страница 33: ...SET MAINTENANCE SET A COM SET MAINTENANCE A Analog transmission E Ethernet EA Ethernet and analog transmission E COM SET MAINTENANCE EA COM SET MAINTENANCE 1 9 M MODE Switch to the regular maintenance...

Страница 34: ...l the reading is stabilized PW AL1 AL2 F LCD 1 1 ZERO Press SET key 1 1 ZERO MAINTENANCE Current Concentration Value Display Press the SET key to perform the zero adjustment 1 1 ppm ZERO SET MAINTENAN...

Страница 35: ...isplay 5 0 ppm AL 1 MAINTENANCE Second Alarm Setpoint Display 10 0 ppm AL 2 MAINTENANCE Alarm Delay Time Display seconds 2 ALM DLY MAINTENANCE Zero suppression value display 0 9 ppm SUPPRESS MAINTENAN...

Страница 36: ...ning off the gas detector the inhibit point skip on the upper central system must be activated Decide whether the power can be turned off by checking the operation of the devices connected to the exte...

Страница 37: ...factory set Although the alarm delay time standard two seconds works in the gas detector to prevent a false activation it can be canceled if not needed Display Operation Gas Concentration Display In c...

Страница 38: ...ttern H HH Alarm Pattern L LL oxygen deficiency alarm Alarm delay time 2 seconds Alarm delay time 2 seconds Alarm lamp ALM 1 Alarm contact terminal plate 5 6 Alarm lamp ALM 2 Alarm contact terminal pl...

Страница 39: ...k that the gas concentration reading is dropped Assuming that gases remain wear protective equipment to avoid dangers and go to the gas leak point and then check if gases remain using a portable gas d...

Страница 40: ...ode gas alarm 4 20 mA concentration output Concentration data Alarm bits 3 Initial Clear Depending on the setting of 4 2 5 mA setting 2 5 mA 4 mA HOLD 4 20 mA setting 4 mA Initial bit 4 Maintenance Mo...

Страница 41: ...a time server can correct the clock at regular intervals Modbus slave function Modbus TCP Works as a Modbus slave and feeds back a setting value in response to a read request or changes a setting val...

Страница 42: ...anagement rules of gas alarm This function is used to hide suppress the fluctuation of the reading under the setting value indicating zero Example COS 0 15 ppm rated range 0 5 ppm COS Suppression Func...

Страница 43: ...f sensor used with the gas detector might have sensitivity variations after being used for a long period This function compensates the deterioration of the gas sensitivity among the sensitivity variat...

Страница 44: ...e than the one that has been replaced CHG UNIT USED SEN Check for correct installation if this message is displayed because an old sensor might have been installed by mistake C 02 CHG SPEC MAINTENANCE...

Страница 45: ...rrect perform the zero adjustment after ensuring that no other gases exist around it Flow rate check See the flow rate indicator to check for abnormalities Filter check Check the dust filter for dust...

Страница 46: ...Inspects the alarm circuit by using the alarm test function Checks the alarm lamps Checks each activation of ALM1 and ALM2 Checks the external alarm Checks the activation of the external alarm Span a...

Страница 47: ...ET 3 MAINTE Regular replacement operation pump stop This is not used on the gas detector SET 4 F TEST Fault alarm test P52 Environmental setting 2 P53 2 10 SETTING2 Functions setting SET 0 ADDRESS Add...

Страница 48: ...7 Mode Item LCD display Details Pyrolyzer data display P62 2 11 PL DATA Display various pyrolyzer data Fault investigation 2 12 FAULT Not used Catalyst aging function P64 2 13 PL AGING Perform aging o...

Страница 49: ...INTENANCE SET 0 2 ppm GAS TEST MAINTENANCE Alternate Display 0 2 ppm COS MAINTENANCE 2 1 ZERO Perform zero adjustment 2 1 ZERO MAINTENANCE SET Zero adjustment P65 2 2 SPAN Perform span adjustment 2 2...

Страница 50: ...TEMP MAINTENANCE 2 8 WARMTIME Display the suggested warm up completion date time for semiconductor type SGU which requires a long warm up time This is not used on the gas detector 2 8 WARMTIME MAINTEN...

Страница 51: ...gate and analyze the causes of faults This is not used by the user 2 12 FAULT MAINTENANCE 2 13 PL AGING Used to perform aging of catalyst 2 13 PL AGING MAINTENANCE SET Catalyst aging function P64 2 14...

Страница 52: ...ered to external circuits even though the gas detector is in the detection mode SET 0 INHIBIT MAINTENANCE SET OFF INHIBIT MAINTENANCE ON INHIBIT MAINTENANCE INHIBIT SET 1 ALM P Set the alarm value SET...

Страница 53: ...kip this menu 10 0ppm AL 2 MAINTENANCE Return to SET 1 ALM P Fault Alarm Test 2 9 SET 4 PW A1 A2 F LCD SET 4 F TEST Press SET key SET 4 F TEST MAINTENANCE Fault Alarm Test ON OFF Select either ON OFF...

Страница 54: ...CE SET 1 DAY TIME Set the date time of the main unit SET 1 DAY TIME MAINTENANCE SET Date and time setting P58 SET 2 SUPPRESS Set the suppression value Change the value by pressing the or key and then...

Страница 55: ...ss the SET key to confirm the selection When ON is selected the external output 4 20 mA is output even during an alarm test When OFF is selected the output is kept at the one just before the alarm tes...

Страница 56: ...ector is used in a normal way because it determines how the gas detector functions Auto reset setting nL SET 10 FLT PTRN MAINTENANCE SET 11 AT FLOW Set the flow rate auto adjustment Select either ON O...

Страница 57: ...electrochemical type ESU is installed Select either ON OFF and then press the SET key to confirm the selection When ON is selected the sensitivity correction function is activated SET 14 S ASSIST MAIN...

Страница 58: ...the selection When ON is selected the backlight lights up continuously When SAVE is selected the backlight is usually turned off but lights up only during an operation or event SET 17 BK LIGHT MAINTEN...

Страница 59: ...and then press the SET key to confirm the value 12 00 2009 01 01 MAINTENANCE Date Setting Change the value by pressing the or key and then press the SET key to confirm the value 12 00 2009 01 01 MAINT...

Страница 60: ...t Alarm Contact Setting Select either nd de energized nE energized and then press the SET key to confirm the selection Press the MODE key to skip this menu nd FLT RLY MAINTENANCE nE FLT RLY MAINTENANC...

Страница 61: ...E MAC2 MAC6 IP Address Setting Change the value by pressing the or key and then press the SET key to confirm the value There are IP addresses from IP1 to IP4 Select one using and keys and set them one...

Страница 62: ...DHCP MAINTENANCE ON DHCP MAINTENANCE Setting result record To record the results of the settings made in the above press the SET key SAVE MAINTENANCE SET SPAN Y N MAINTENANCE Return to MAC Address Che...

Страница 63: ...CE PL 2 PL SET Press the or key to change the setting of pyrolyzer heater temperature and then press the SET key to confirm the value PL 2 PL SET MAINTENANCE SET 120 C PL SET MAINTENANCE PL 3 PL AMP1...

Страница 64: ...3 3V MAINTENANCE PL 7 PL 5 0V Display the pyrolyzer 5 0 V output PL 7 PL 5 0V MAINTENANCE SET 5000 mv PL 5 0V MAINTENANCE PL 8 PL VER Display the pyrolyzer version PL 8 PL VER MAINTENANCE SET 01234 A...

Страница 65: ...he pyrolyzer temperature is adjusted to 700 C automatically When 700 C is reached it is kept for two hours 700 C PL AGING MAINTENANCE Two hours later the pyrolyzer temperature is adjusted to the setti...

Страница 66: ...d PW A1 A2 F LCD 2 1 ZERO Press SET key 2 1 ZERO MAINTENANCE Current Concentration Value Display Press the SET key to perform the zero adjustment 1 1 ppm ZERO SET MAINTENANCE Zero Adjustment Completed...

Страница 67: ...se of less than 10 FS this is not switched to the next menu 2 0 ppm SPAN GAS MAINTENANCE Span Adjustment Because the reading will be fixed adjust the introduced gas concentration by pressing the or ke...

Страница 68: ...e menu returns to 2 2 SPAN automatically 8 0 ppm SPAN END MAINTENANCE Return to 2 2 SPAN When the span adjustment fails 8 0 ppm SPAN NG MAINTENANCE Return to 2 2 SPAN CAUTION Exhaust gas must be colle...

Страница 69: ...sed or decreased by pressing the or key the flow rate changes accordingly Adjust the flow rate to 0 5 L min and then press the SET key Press the MODE key to cancel this menu 50 0 50 L M MAINTENANCE Cl...

Страница 70: ...nit 0 5 years 1 2 years 1 2 Flow sensor 1 year 5 years 1 4 Fan 0 5 years 2 4 years 1 5 Catalyst tube 0 5 years 1 year 1 6 Sensor unit 0 5 years 1 year 1 7 Tubes and joints 0 5 years 3 8 years 8 C4F6 C...

Страница 71: ...st be checked by a qualified service engineer Since the catalyst tube has an influence over gas sensitivity calibration needs to be performed after replacement Please contact RIKEN KEIKI Sensor Unit R...

Страница 72: ...light In a place where gases solvents or vapors are not present 8 2 Procedures to relocate the gas detector or use it again When the gas detector is relocated select a new place in accordance with 4 2...

Страница 73: ...or turn OFF the power and contact RIKEN KEIKI immediately When the gas detector is disposed of it must be treated properly as an industrial waste in accordance with the local regulations WARNING Do no...

Страница 74: ...cuit Check the wiring of gas detector and related devices around it Fuse disconnection Find out why the fuse has blown and take appropriate actions before replacing it If the power still cannot be tur...

Страница 75: ...ilter Replace the dust filter Unstable flow caused by bended or clogged suction tube or exhaust tube Fix the defective parts Pressure difference is present in the sampling condition The flow rate is e...

Страница 76: ...of the clock inside the gas detector Make a setting of Date Time Note that when the sensitivity correction function of ESU is used correction may not be made properly If such a symptom is observed re...

Страница 77: ...information on actions such as removal filter please contact RIKEN KEIKI Disturbance by noise Turn off and restart the gas detector If such a symptom is observed frequently take appropriate measures...

Страница 78: ...play First ALM1 lamp blinks or lights up Red Second ALM2 lamp blinks or lights up Red Gas alarm pattern Auto reset or self latching Gas alarm contact No voltage contact 1a or 1b 2 step independent De...

Страница 79: ...ication Digital transmission Ethernet Analog transmission 4 20 mA DC no insulation load resistance under 300 Transmission cable Shielded twisted pair cable 1 25 sq of KPEV S etc 1P Digital transmissio...

Страница 80: ...electrochemical type sensor The electric potential between the working electrode and reference electrode is kept at a certain level by a potentiostat circuit The gas to be detected is electrolyzed di...

Страница 81: ...the analog signal output must be used for an indicator or external recorder If these outputs are used to control other units we shall not be responsible for any malfunctions 4 Because the contact poin...

Страница 82: ...nd a signal to indicate the maintenance mode status is sent out to the external output signal As a result maintenance can be performed on a single unit of the gas detector Initial clear The reading is...

Страница 83: ...11 Definition of Terms 82...

Отзывы: