7. Maintenance
7-1. Maintenance intervals and items
- 44 -
7
7.
Maintenance
The gas detector is an important instrument for the purpose of safety.
To maintain the performance of the gas detector and improve the reliability of safety, perform a regular
maintenance.
7-1. Maintenance intervals and items
Daily maintenance: Perform maintenance before beginning to work.
Monthly maintenance: Perform maintenance on the alarm circuit (alarm test) once a month.
Regular maintenance: Perform maintenance once or more for every six months to maintain the
performance as a safety unit.
Maintenance
item
Maintenance content
Daily
mainte-
nance
Monthly
mainte-
nance
Regular
mainte-
nance
Power supply
check
Check that the power lamp lights up.
○
○
○
Concentration
display check
Check that the concentration display value is zero.
When the reading is incorrect, perform the zero
adjustment after ensuring that no other gases exist
around it.
○
○
○
Flow rate check
See the flow rate indicator to check for abnormalities.
○
○
○
Filter check
Check the dust filter for dust or clogging.
○
○
○
Alarm test
Inspect the alarm circuit by using the alarm test
function.
○
○
Span adjustment Perform the span adjustment by using the calibration
gas.
○
Gas alarm check Check the gas alarm by using the calibration gas.
○
<About Maintenance Services>
We provide services on regular maintenance including span adjustment, other adjustments and
maintenance.
To make the calibration gas, dedicated tools, such as a gas cylinder of the specified concentration and
gas sampling bag must be used.
Our qualified service engineers have expertise, knowledge and other information on the dedicated tools
used for services, along with other products. To maintain the safety operation of the gas detector, please
use our maintenance service.
Typical maintenance services are listed as follows. Please contact RIKEN KEIKI for more information.