7. Maintenance
7-2. Maintenance (regular maintenance) mode
38
<Span Adjustment>
2-2.SPAN
Perform the span
adjustment.
Press the SET button.
Press the SET button
again.
Supply calibration gas
and press the SET
button.
"PASS" is displayed
when adjustment is
completed, and "FAIL"
is displayed when it
fails.
If "FAIL" is displayed,
perform span
adjustment again.
Span adjustment will
end.