9. Troubleshooting
9. Troubleshooting
46
9
9.
Troubleshooting
The Troubleshooting does not explain the causes of all the malfunctions which occur on the monitor. This
simply helps to find the causes of malfunctions which frequently occur.
If the monitor shows a symptom which is not explained in this manual, or still has malfunctions even though
remedial actions are taken, please contact RIKEN KEIKI.
Symptom/Display
Causes
Actions
The power cannot
be turned on.
The power switch is
turned off.
Turn on the power switch.
Abnormalities/momentary
blackout of power supply
system
Provide the rated voltage.
Take measures such as checking or adding the
UPS, power supply line filter and insulation
transformer.
The main unit is not
installed properly.
Check whether the main unit is properly attached to
the wall-mounted unit.
Cable abnormalities
(open circuit/not
connected/short circuit)
Check the wiring of the monitor and related devices
around it.
Abnormal
operations
Disturbances by sudden
surge noise, etc.
Turn off and restart the monitor.
If a symptom like this is observed frequently, take
appropriate measures to eliminate the noise.
Sensor
abnormalities
E-1
The remote sensor is not
connected or improperly
connected.
Check that the remote sensor is connected and the
connectors of the sensor board are securely
fastened.
Errors in communication
with the remote sensor
Replace the sensor board with new one.
Clock
abnormalities
E-9
Abnormalities of the clock
inside the monitor
Make a setting of date/time. If a symptom like this is
observed repeatedly, the built-in clock is seemingly
malfunctioning. Thus, it must be replaced. Please
contact RIKEN KEIKI.
System
abnormalities
E-9 SYSTEM
The rated voltage is not
supplied to the monitor.
Check the power supply, and supply the rated
voltage.
Abnormalities of ROM,
RAM or EEPROM inside
the monitor
Please contact RIKEN KEIKI.
The reading drops
and it remains so.
Drifting of sensor output
Perform the zero calibration.
Presence of interference
gas
Disturbances by interference gases, such as
solvents, cannot be eliminated completely. For
information on actions, such as removal filter,
please contact RIKEN KEIKI.
Environmental changes
Perform the zero calibration.
Содержание GM-600
Страница 1: ...PT2E 2350 E O Gas Monitor GM 600 Operating Manual...
Страница 36: ...7 Maintenance 7 2 Maintenance regular maintenance mode 36 2 03 LCD Not used 2 04 Not used To 2 00...
Страница 41: ...7 Maintenance 7 2 Maintenance regular maintenance mode 41 2 4A DATE Not used...
Страница 44: ...7 Maintenance 7 2 Maintenance regular maintenance mode 44 2 4J TYPE Not used 2 4L DISP Not used...
Страница 49: ...10 Product Specifications 10 Product Specifications 49 10 3 Optional accessories Installation board 1 pc...
Страница 52: ......