6. Alarm Activation and Functions
6-1. Gas alarm activation
28
<Alarm Display (ALM1: Yellow Backlight), (ALM2: Red Backlight)>
The alarm consists of two steps. The yellow or red backlight lights up when the respective alarm setpoint
value is reached to or exceeded.
Contact activation
The contact is activated when the gas concentration reaches or exceeds the alarm setpoint value. The
contact activation is reset automatically when the gas concentration drops below the alarm setpoint value.
Response to gas alarm (In case of responding to leaked gas)
When a gas alarm is triggered, take actions in accordance with your management rules of gas alarm.
Normally, take the following actions.
- Check the reading of the monitor.
- If the gas alarm display continues to be displayed, close the main valve of the gas, and then check
that the gas concentration reading is back to normal.
NOTE
If a gas leak is momentary, the reading may already be back to normal when checking it. In addition,
when the alarm is triggered by noise or other incidental conditions other than a gas, the reading may
have already returned to normal.
WARNING
If the MODE or SET button is pressed to enter another mode while an alarm is triggered, the
contact activation is reset.
CAUTION
If the monitor enters maintenance mode while a gas alarm is triggered, and if it returns to the
detection mode while gas still exists, it does not trigger any alarms for about 15 minutes and cannot
check if gases exist.
Alarm lamp (ALM1)
First alarm setpoint
"Alarm pattern (H-HH)"
Second alarm setpoint
Alarm lamp (ALM2)
Alarm contact (Terminal plate 3, 4)
Concentration
Alarm delay time (2 seconds)
Time
Alarm delay time (2 seconds)
Содержание GM-600
Страница 1: ...PT2E 2350 E O Gas Monitor GM 600 Operating Manual...
Страница 36: ...7 Maintenance 7 2 Maintenance regular maintenance mode 36 2 03 LCD Not used 2 04 Not used To 2 00...
Страница 41: ...7 Maintenance 7 2 Maintenance regular maintenance mode 41 2 4A DATE Not used...
Страница 44: ...7 Maintenance 7 2 Maintenance regular maintenance mode 44 2 4J TYPE Not used 2 4L DISP Not used...
Страница 49: ...10 Product Specifications 10 Product Specifications 49 10 3 Optional accessories Installation board 1 pc...
Страница 52: ......