21
8-1.
Standard specifications
Model
FI-800
Measuring principle
Fourier analysis, combined wave length optical interferometric
method.
Measuring method
Sample drawing type by external sampling unit.
Response time
Within 30 sec. to get 90% of concentration (In case of 300±
25ml/min gas supply)
Analog outputs
4~20mADC(Current throw type), into 300Ω impedance max.
Concentration display Digital LCD display
Alarm outputs
Relay contact (a-contact for each AL1 and AL2), rated 1A at
125VAC/30VDC (Resistive load).
Self-diagnostic
function
Low sample flow, decrease of light amount for interferometric
light source, decrease of interferometer contrast.
Power requirement
100VAC±10%~220VAC±10%, 50/60Hz, max 8VA
Operating temp. &
humidity
-10℃~+40℃, below 80%RH(Condensing gas inside FI-800 is
not possible)
Outer dimensions
Approx. 220(W) x 332(H) x 122(D)mm
Weight
Approx 16Kg.
Structure
Flame-proof ExdⅡB+H2T4
* Measuring gas and range are shown in attached 「Specifications for measuring gas」.
8-2.
Standard accessories
・
Control key (1 pce.)
・
Specifications for measuring gas(1 sheet)
・
Operating instruction manual(1 sheet)
・
Installation work manual(1 sheet)
8.Product specifications