8
Flashing CONT light
This is a cautional indication for the decrease
of contrast. This means that the interferometric
sensor is being dirty. Recommend to replace or
repair it before interruption of measurement.
2-6.
Trouble indications and countermeasures
When the abnormality is detected during
selfdiagnosis or measurement, PW/TR
light flashes and relay contact for
trouble alarm works.
In this occasion, error code number is
displayed and outputs for 4-20mA is
fixed at 0.5mA.
Make necessary countermeasure if error code number is displayed.
Contents of error code and countermeasures
Error code
Contents and countermeasures
E-00
This means the abnormality of the system.
Replacement/repair of main PC board is required
E-01
This means the interruption of measurement due to the
deterioration of inteferometric sensor.
Flashing of BRIGHT light means the decrease of light
amount for light source remarkably and flashing CONT
light means the decrease of contrast remarkably.
In any case, replacement/repair of interferometric
sensor is required.
E-02
This means the possibility of damage for setting data of
measuring gas or reference gas, a damage for alarm level
or sensor sensitivity data being set at maintenance
mode. Re-setting for various data or replacement/repair
of main PC board is required.
E-05
This means that the sample flow rate for sample gas or
reference gas to be supplied from GAS IN or REF IN is
zero.
It is thought that external pump, aspirator or reducing
valve is defective or pipings are blocked.
Flashing
Display at abnormality
Display of error code number
1 2
Flashing