14
3-8.
Setting of AUTO REF CAL CYCLE(Option)
Pressing SET switch during is displayed, auto reference calibration
cycle is displayed. This is an optional function. The FI-800 which has no optional
function is set to 「 ACC.0 」.
3-9.
Display of ROM number
Pressing SET switch during is displayed, program number
(our control No.) and sum level in FI-800 are shown one after the other.
This function is to confirm by field service engineer and is not important
function for user.
Pressing ●mark of MAINTENANCE switch, it returns to .
3-10.
Display of specification number
Pressing SET switch during is displayed, specification number
(our control No.) of FI-800 is displayed.
This function is to confirm by field service engineer and is not important function
for user.
Pressing ●mark of MAINTENANCE switch, it returns to .
3-11.
Pressure/temperature display in sensor part.
Pressing SET switch during or is displayed, gas
pressure (kPa, Absolute pressure) or temperature (℃) around sensor part is
displayed.
A C C. 0
Select 0~7
with ▲▼switch
and enter with
SET switch
SET switch
A C C. 0
It is required to make exclusive piping and wiring construction to use the
function of auto reference calibration.
Be sure to set to 「ACC.0」 if this function is not used. If incorrect setting
is made, zero level goes down minus side and normal measurement cannot
be performed
!
WARNING