Wheel diameter measuring gauge, IDK-BT
IDK-BT [Revision 2.2] 4
th
September 2017
12
RF510.11.000
Figure 4
If readings of the sensor are sufficiently stable, press the
Blue
button. Sensor
zero position is calculated in the device coordinate system, and prompt ap-
pears to save calibration results. Press the
Green
or
Red
button to confirm or
cancel saving of the results, respectively.
11.4. Calibration of the device base
Device base calibration mode being enabled, the display shows the value of
reference diameter used for calibration.
If editing of the diameter value is not required, go to the next step. To edit the
diameter value, press the
Blue
button, and the digit to be edited starts blink-
ing. Changing over between the digits is made by pressing the
Blue
button
while changing of values is made by pressing the
Green
button. When editing
is finished, press the
Red
button and confirm or cancel saving of the parame-
ter by pressing the
Green
button or
Red
button, respectively.
Press the
Green
button, and the reference diameter value starts blinking,
which means that the device must be placed onto a gage with the diameter
value set at the previous step. The device supports must be firmly forced
against the reference block.
Press the
Green
button, and the display shows current reading of the sensor
in the device coordinate system. By moving the device, assure that repeatabil-
ity of measurement results is obtained.
If readings of the instrument are sufficiently stable, press the
Blue
button. De-
vice base value is calculated, and prompt appears to save calibration results.
Press the
Green
or
Red
button to confirm or cancel saving of the results, re-
spectively.