11. Technical Specifications
Operating conditions:
10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
Storage conditions:
-20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight:
510g (including batteries)
Dimensions:
170 x 135x 41 mm
Measuring procedure:
oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range:
60 - 255 mmHg – systolic blood pressure
30 - 200 mmHg – diastolic blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
NiMH Battery pack, 4.8V 2400mAh
Mains adapter DC 7.5V, 1.5 mA
Battery lifetime:
approx. 1000 measurements (after the battery
pack is fully charged)
IP Class:
IP20
Reference to standards:
IEC 60601-1;;
IEC 60601-1-2 (EMC)
DIN EN ISO 81060-1
IEC 80601-2-30
EN 1060-1 /-3
12. EMC
Guidance and manufacture’s declaration – electromagnetic emission
The RBP-100 Blood Pressure Monitor and the RBP-100 USB Blood Pressure Monitor is intended
for use in the electromagnetic environment specified below. The customer of the user of the RBP-
100 Blood Pressure Monitor and the RBP-100 USB Blood Pressure Monitor should assure that it
is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The RBP-100 Blood Pressure Monitor and the
RBP-100 USB Blood Pressure Monitor RF
energy only for its internal function. Therefore,
its RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Class B
The RBP-100 Blood Pressure Monitor and the
RBP-100 USB Blood Pressure Monitor is
suitable for use in all establishments, other
than domestic and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Pass
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Pass
Seite 19 von 23
11. Technical Specifications
Operating conditions:
10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
Storage conditions:
-20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight:
510g (including batteries)
Dimensions:
170 x 135x 41 mm
Measuring procedure:
oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range:
60 - 255 mmHg – systolic blood pressure
30 - 200 mmHg – diastolic blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
NiMH Battery pack, 4.8V 2400mAh
Mains adapter DC 7.5V, 1.5 mA
Battery lifetime:
approx. 1000 measurements (after the battery
pack is fully charged)
IP Class:
IP20
Reference to standards:
IEC 60601-1;;
IEC 60601-1-2 (EMC)
DIN EN ISO 81060-1
IEC 80601-2-30
EN 1060-1 /-3
12. EMC
Guidance and manufacture’s declaration – electromagnetic emission
The RBP-100 Blood Pressure Monitor and the RBP-100 USB Blood Pressure Monitor is intended
for use in the electromagnetic environment specified below. The customer of the user of the RBP-
100 Blood Pressure Monitor and the RBP-100 USB Blood Pressure Monitor should assure that it
is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The RBP-100 Blood Pressure Monitor and the
RBP-100 USB Blood Pressure Monitor RF
energy only for its internal function. Therefore,
its RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Class B
The RBP-100 Blood Pressure Monitor and the
RBP-100 USB Blood Pressure Monitor is
suitable for use in all establishments, other
than domestic and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Pass
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Pass
Содержание 04045396208797
Страница 1: ...RBP 100 Bedienungsanleitung DIGITALES BLUTDRUCKMESSGERÄT 0124 ...
Страница 31: ...RBP 100 Instruction for use DIGITAL BLOOD PRESSURE MONITOR 0124 ...
Страница 61: ...RBP 100 Instructions d utilisation TENSIOMÈTRE NUMÉRIQUE 0124 ...
Страница 91: ...Instrucciones de uso MONITOR DE PRESIÓN ARTERIAL DIGITAL 0124 RBP 100 ...
Страница 121: ...RBP 100 Инструкция по эксплуатации ЦИФРОВОЙ ТОНОМЕТР 0124 ...
Страница 151: ...RBP 100 Istruzioni per l uso MISURATORE DIGITALE DELLA PRESSIONE SANGUIGNA 0124 ...
Страница 181: ...RBP 100 ...
Страница 182: ...RBP 100 ...
Страница 183: ...RBP 100 ...