8. ERROR MESSAGES
If an error occurs during the measurement, the measurement is interrupted and an error
message, e.g. «ERR 3», is displayed.
Note:
- When an error occurred during the measurement, the reading is not stored in the memory.
Seite 15 von 24
Error
Description
Potential cause and remedy
«
ERR 1
»
Signal to weak
The pulse signals on the cuff are too weak. Re-position
the cuff and repeat the measurement.
«
ERR 2
»
Error signal
During the measurement, error signals were detected by
the cuff, caused for instance by movement or muscle
tension. Repeat the measurement, pay attention the
patient is keeping his arm still.
«
ERR 3
»
Abnormal cuff
pressure / inflation /
deflation
An adequate pressure cannot be generated in the cuff.
A leak may have occurred. Check that the cuff is
correctly connected and is not too loose. Replace the
batteries if necessary. Repeat the measurement.
«
ERR 5
»
Abnormal result
The measuring signals are abnormal and no result can
therefore be displayed. Read through the checklist for
performing reliable measurements and then repeat the
measurement.
«
HI
»
Pulse or cuff Pressure
too high
The pressure in the cuff is too high (over 299 mmHg)
OR the pulse is too high (over 200 beats per minute).
The patient needs to relax for 5 minutes and then repeat
the measurement.
Note:
-
Irregular heartbeat detection is inactivated
when pulse or pressure are lying above these
limits.
«
LO
»
Pulse too low
The pulse is too low (less than 40 beats per minute).
Repeat the measurement.
Note:
-
Irregular heartbeat detection is inactivated
when pulse is below this limit.
Note:
-
When an error occurred during the measurement, the reading is not stored in the memory.
9.
Safety, Care, Accuracy Test and Disposal
9.1 Safety and protection
9.1.1 Follow instructions for use. This document provides important product operation and safety
information regarding this device. Please read this document thoroughly before using the
device and keep for future reference.
9.1.2 This device may only be used for the purposes described in these instructions. The
manufacturer cannot be held liable for damage caused by incorrect application.
9.1.3 This device comprises sensitive components and must be treated with caution. Observe the
storage and operating conditions described in the «Technical Specifications» section.
9.1.4 Protect it from:
- water and moisture
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
9.1.5 The cuffs are sensitive and must be handled with care.
Содержание 04045396208797
Страница 1: ...RBP 100 Bedienungsanleitung DIGITALES BLUTDRUCKMESSGERÄT 0124 ...
Страница 31: ...RBP 100 Instruction for use DIGITAL BLOOD PRESSURE MONITOR 0124 ...
Страница 61: ...RBP 100 Instructions d utilisation TENSIOMÈTRE NUMÉRIQUE 0124 ...
Страница 91: ...Instrucciones de uso MONITOR DE PRESIÓN ARTERIAL DIGITAL 0124 RBP 100 ...
Страница 121: ...RBP 100 Инструкция по эксплуатации ЦИФРОВОЙ ТОНОМЕТР 0124 ...
Страница 151: ...RBP 100 Istruzioni per l uso MISURATORE DIGITALE DELLA PRESSIONE SANGUIGNA 0124 ...
Страница 181: ...RBP 100 ...
Страница 182: ...RBP 100 ...
Страница 183: ...RBP 100 ...