4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG
Page 21
Presentation of the results on the printout after a successful measurement:
V2.2a dd/mm/yy D
LAB.: RIELE BERLIN
USER 1: M.MUSTERMANN
DATE: 07/17/17
TIME: 08:44:12
METHOD 11: KIN/F/Rb
PROGRAM: 11
FACTOR: 1.000
WAVELENGTH: 340nm
TEMPERATURE: 37C
DELAY: 60s
DELTAS: 5
TIME/DELTA: 18s
UNIT: U/l
- - - - - - - - - - - -
MEASURE ZERO
OLD Rb[A]: 0.000
NO. ABS. RESULT
1 0.107 150.8
R^2: 0.9994
NO. TIME [s] ABS.
NO. TIME [s] ABS.
1 0 0.718
2 30 0.734
3 60 0.750
4 90 0.767
5 120 0.785
6 150 0.805
7 180 0.825
ABS.1: 0.642
ABS.2: 0.750
1: 0.0312
2: 0.0320
3: x 0.0345
4: x 0.0364
5: x 0.0387
6: 0.0417
DELTA REG: 0.0365
ABS = ABS.2 - ABS.1
RESULT = DELTA REG x FACTOR
Printout after confirming [DIAGR.]
and [PRINT]
The chronic of the kinetic is printed.
The axis of time is marked by [s]; the
axis of extinction is marked by [ABS].
In the next step a chart including all
data points is printed.
Printout after confirming [MODE],
[MODE] and [DETAIL].
ABS/min
In case of an improvement of the
linearity considering the three best
deltas these are marked with a “x”. If
this procedure does not lead to an
improvement
all
deltas
are
reconsidered in the calculation
procedure and none of the deltas are
marked.