4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG
Page 8
2 - INSTALLATION
2
INSTALLATION
2.1
DELIVERY
Check the device and contents of the enclosed box as follows on visible transport damages and completeness:
1
Operator’s Manual
1
Dust cover
2
Fuses for line power
1
Mains cable
2
Thermal printer paper
1
Top cover small for printer
Inform the sales office immediately about transport damages. Keep the original packaging for a possible
return.
2.2
PREPARATION FOR INSTALLATION
Place the device on a stable, level surface. Do not obstruct the input air at the bottom and the waste air at the
back plate to guarantee the ventilation of the device.
If the device was exposed to extraordinary fluctuation in temperature and/or humidity, it must acclimatize suffi-
ciently before operation.
2.3
INSTALLATION
Photometer
4040
+
operates at any line voltage between 100 V
AC
and 240 V
AC
at 50/60 Hz. The device plug of the
mains cable must be put into the socket at the back of the device and the mains plug into a grounded mains
socket.
While connecting or disconnecting an external device (PC, printer) to Photometer
4040
+
both devices must
be switched off.
Switch on Photometer
4040
+
by the mains switch at the back.
Greeting screen:
After switching on copyright, website, type of
device
and
version
designation
are
displayed and - in the case of activated
printer - printed out.
( c ) R O B E R T R I E L E
G m b H & C o K G
P H O T O M E T E R 4 0 4 0 +
V 2 . 2 a d d / m m / y y D
w w w . r i e l e . d e