4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG
Page 34
6.4.9
Calculation procedure 9 (FTK/F/Rb)
Method at which a reagent blank is measured after an incubation time (
A
RB,0
) and after a reaction time (
A
RB,1
) and also a sample after an incubation time (
A
S,0
) and after a reaction time (
A
S,1
).
The difference from the change of the test and the change of the reagent blank is multiplied by a predefined factor
F. The reagent blank A
RB
is entered or measured once.
During the procedure the dialog asks for the use of a reagent blank. The default value is OFF. To continue with-
out reagent blank press [NO].
After each measurement the next sample can be measured with [NEXT].
With [RESULT] it is possible to measure the same sample again.
Calculation procedure ....................................................... CP 9
Characteristic ........................................................ FTK / F / Rb
Method ................................................. Fixed Time with Factor
Calculation formula ......... C = F
( |A
S,0
- A
S,1
| - |A
RB,0
- A
RB,1
| )
Factor ............................................................... given / entering
Reagent blank ........................................ entering or measuring
V2.2a dd/mm/yy D
LAB.: RIELE BERLIN
USER 1: M.MUSTERMANN
DATE: 07/17/17
TIME: 08:44:12
METHOD 29: CK-MB
PROGRAM: 9
FACTOR: 2751.3
WAVELENGTH: 340nm
TEMPERATURE: 37C
INCUBATION: 120s
REACTION: 180s
MAX.UNITS: 1500
UNIT: U/l
- - - - - - - - - - - -
MEASURE ZERO
Rb[A]: 0.000
NO. ABS. RESULT
1 1.005 910.7
DELTA [A]: 0.331
2 1.029 1128.1
DELTA [A]: 0.410
3 0.829 1381.2
DELTA [A]: 0.502
Start method selection in the main menu.
See chapter:
4.1 - Measurement with programmed methods
4.2 - Measurement with basic methods
In the case of activated printer the print-out of the method data
follows.
The measuring window is shown.
Method procedure:
Insert / measure zero solution
Without reagent blank (insert / measure optionally)
Insert / measure sample
Insert / measure sample
Insert / measure sample