4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG
Page 26
6.4.1
Calculation procedure 1 (C/F)
Method at which a measured sample value A
S
is multiplied with a predefined factor F.
Calculation procedure ....................................................... CP 1
Characteristic .................................................................... C / F
Method .................................................... End Point with Factor
Calculation formula .................................................. C = F
A
S
Factor ............................................................... given / entering
PHOTOMETER 4040+ # 1001
V2.2a dd/mm/yy D
LAB.: RIELE BERLIN
USER 1: M.MUSTERMANN
DATE: 07/17/17
TIME: 08:44:12
METHOD 20: HEMOGLOBIN
PROGRAM: 1
FACTOR: 29.4
WAVELENGTH: 405nm
TEMPERATURE: 37C
DELAY: 5s
MAX.UNITS: 25
UNIT: g/l
- - - - - - - - - - - -
MEASURE ZERO
NO. ABS. RESULT
1 0.675 19.8
2 0.843 24.8
Start method selection in the main menu.
See chapter:
4.1 - Measurement with programmed methods
4.2 - Measurement with basic methods
In the case of activated printer the print-out of the method data
follows.
The measuring window is shown.
Method procedure:
Insert / measure zero solution
Insert / measure sample
Insert / measure sample