4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG
Page 13
4.2
Measurement with basic methods
A photometric test can be executed by a method already permanently programmed, but open in all setting
parameters. 14 different methods with different calculation procedures are available. Each of these methods can
serve as prototype for a method programmed by the operator.
Available are:
Absorbance measurement
Concentration measurement / end point
measurement
Fixed time kinetic / two point kinetic
Kinetic
Transmission
Scrolling through all methods is possible by
[PAGE]. The current page is shown at the
right upper screen corner. By [END] the
program returns to the main menu.
A method is selected by pressing the corre-
sponding key.
The following abbreviations are used for the
distinction of the methods:
CONC. = concentration
KIN = kinetic
FTK = fixed time kinetic
F = factor
STD = standard
RB = reagent blank
SB = sample blank
Further information:
Chapter: 6 - CALCULATION
PROCEDURES
4.3
Method editor
Each photometric test can be permanently
stored with its setting parameters by the
method editor.
With the functions of the method editor are
possible the new installation, the change
and removing a method.
By [LIST] an overview of the programmed
methods can be printed and transmitted via
the serial interface.
Further information:
Chapter: 7 - METHOD EDITOR
B A S I C M E T H O D S P A G E 1 / 4
C O N C . W . F A C T O R P A G E
C O N C . W . F A C T O R R B
C O N C . W . F A C T O R S B
C O N C . W . F A C T O R R B S B E X I T
M E T H O D N E W / C H A N G E / C O P Y
M E T H O D C O P Y L I S T
M E T H O D E D I T
M E T H O D N E W
M E T H O D D E L E T E E X I T
o 3 7 . 0 ° C i 3 7 . 0 ° C 0 7 / 2 4 / 2 0 1 7 1 0 : 1 4