MSI-4260M Port-A-Weigh Crane Scale
30
Visit our website
7.0
Troubleshooting/Maintenance
7.1 Troubleshooting
Problem
Possible Cause
Solution
Display is blank when the POWER button is
depressed
Discharged batteries
Replace batteries
Defective battery
Replace batteries
Corroded battery or battery contacts Clean the battery contacts
Defective switch or circuit board
Requires authorized service
Display does not function properly, the front
panel button does not function normally or
the scale will not turn off
Improperly updated software
Reinstall the software
Faulty circuit board
Requires authorized service
Loose connectors
Requires authorized service
Scale does not respond to weight changes Out of calibration
Calibrate the unit
Faulty load cell
Replace the load cell
Load cell connector
Check the connector and wires
The display over ranges below 100%
capacity
Tared weight is added to load to
determine overload point
Return to gross weight mode
Zero requires adjustment
Rezero the scale
Too much weight has been zeroed Rezero the scale
The display drifts
AZM (Auto0) is turned off
Turn AZM on
Rapid temperature change,
moving scale outdoors from indoors
Wait until the scale temperature has stabilized
The displayed weight shows a large error Scale not zeroed before load is lifted Zero the scale with no load attached
lb/kg units causing confusion
Select the proper units
Requires recalibration
Recalibrate the unit
Display reading is not stable
Excessive vibration in crane system Increase filtering or increase ‘d’ in Cal
Excessive side loading
Improve load train symmetry
Load cell faulty
Check the load cell connections
Display toggles between
and
Weight exceeds capacity
Reduce weight immediately
Faulty load cell or wiring
Check load cell and load cell wiring
Display toggles between
and
Weight in below the zero range
If the scale is in compression, remove the source
Calibration faulty
Recalibrate
Faulty load cell or wiring
Check the load cell connections
Display toggles between
and
A/D is saturated negative
Check the load cell and load cell wiring
Display toggles between
and
A key is stuck or is being held down Check switches for damage
Ensure that a remote is not transmitting continuously
RF remote device does not work
Units are not paired
Contact Rice Lake Weighing Systems
Unit turns on then immediately turns off
Batteries are low
Replace batteries
Weight will not zero
The system not stable
The stable annunciator must turn on for Zero to function;
Increase the filtering for more stability
Increase the filtering for more stability
Zero is out of range
Legal for Trade units have limited zero range;
Reduce the weight or use Tare instead
The weight will not Zero, Tare or Total
The system is not stable
Wait for Stable annunciator to turn on, or if in a mechanically noisy
crane, increase the filtering or increase the size of the scale increment
‘d’; It is also possible to increase the motion window; Contact Rice
Lake Weighing Systems if you have a problem getting the MSI-4260M
to zero, tare, or total due to stability issues
Setpoint lights blink
Setpoint is enabled and the trigger
point has been reached
Disable set points if they are not needed
Manual total does not work
A function key is not set to “Total”
Set up Func1 or Func2 for “Total”
The weight must be stable
Increase filtering for more stability
Auto Total does not work
The weight must be stable
Wait for stability annunciator to turn on, or increase filtering
Weight thresholds not reached
Capacity must exceed 1% for autototal to work; Capacity must drop
below 0.5% of capacity for additional weighments to register
Table 7-1. Troubleshooting