Configuration
15
3.2.1
Configuration Menu
Figure 3-5. Configuration Menu
CONFIG Menu
Parameter
Choices
Description
Level 2 submenus
GRADS
10000
number
Graduations. Specifies the number of full scale graduations. The value entered must be in
the range 1–60 000 and should be consistent with legal requirements and environmental
limits on system resolution.
To calculate GRADS, use the formula, GRADS =
Capacity
/
Display Divisions
.
Display divisions for primary and secondary units are specified on the FORMAT menu.
ZTRKBN
OFF
0.5D
1D
3D
Zero track band. Automatically zeroes the scale when within the range specified, as long
as the input is within the ZRANGE and scale is at standstill. Selections are ± display
divisions. Maximum legal value varies depending on local regulations.
ZRANGE
1.9%
100%
Zero range. Selects the range within which the scale can be zeroed. The 1.9% selection is
± 1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be at
standstill to zero the scale. Use 1.9% for legal-for-trade applications.
Table 3-2. Configuration Menu Parameters
XXXXXXX
XXXXXXX
XXXXXXX
XXXXXXX
GRADS
10000
OFF
1D
0.5D
3D
ZTRKBN
ZRANGE
1.9%
100%
OFF
2D
1D
3D
MOTBAN
10D
5D
20D
FS+2%
FS+9D
FS+1D
FS
OVRLOA
1
4
2
8
DIGFL1
32
16
64
number
1
4
2
8
DIGFL2
32
16
64
1
4
2
8
DIGFL3
32
16
64
BOTH
PBTARE
NOTARE
KEYED
TAREFN
VERS
ALGOUT
DIG IN
PROGRM
PFORMT
SERIAL
CALIBR
CONFIG
FORMAT
8OUT
32OUT
16OUT
64OUT
DFSENS
DFTHRH
2OUT
128OUT
4OUT
2DD
NONE
5DD
20DD
10DD
50DD
200DD
100DD
250DD
Содержание IQ plus 350
Страница 1: ...43544 IQ plus 350 Digital Weight Indicator Version 1 0 Installation Manual ...
Страница 2: ......