Rh eo n i k M esst ech n i k
RE V . L O ct o b er 2014
.
RHEONIK Coriolis Flowmeter
Operation Manual RHE 07, 08, 11
Page
26
There are additional settings for special operating conditions like:
- FiltArray:
A digital low pass filter for phase shift measurement can be configured. The Filt Array
number is equal to number of filtered measurement cycles. Shortest measurement cycle
time is two sensor oscillations. This filter is very useful for applications with pulsating
flow rates (piston pumps).
Tim e
Mass Flow
act ual Flow ( m ono Pist on Pum p)
m ean Flow
indicat ed Flow
( high Dam ping)
- TFlow:
Digital damping value for display and analog outputs (flow rate).
TFlow
is response time
in X.XX seconds.
- FiltBand:
Flow rate filter band in percentage of half nominal sensor flow rate. Depending on soft-
ware verion, slight and insignificant deviations of this flow rate may be given. Response
time outside filter band is shortest. Inside of this, the adjusted response time TFlow is
active.
Tim e
Mass Flow
act ual Flow
Tim e
Mass Flow
indicat ed Flow
Filt Band