Rh eo n i k M esst ech n i k
RE V . L O ct o b er 2014
.
RHEONIK Coriolis Flowmeter
Operation Manual RHE 07, 08, 11
Page
22
No t e:
Care should be taken to ensure that the flow has been completely stopped before using
the input for zeroing.
- Using the
Reset
function, the input can be used to start a batching process in combina-
tion with totalizer limit outputs.
- Without installed I/O board inside the transmitter RHM the input function has to be
programmed as
Not used
.
2.1.3.6.4. Contact outputs
The output hardware is the same as for frequency output (see section 2.1.3.6.1.).
All outputs can be programmed as follows:
- Limit Flow, Temp or Dens:
Flow rate, temperature or density limit. The output is active below the adjusted setting.
- Limit Mass:
Mass totalizer limit. The output is active below the adjusted totalizer value.
- Error:
Output is activated when a malfunction in the flowmeter occurs or is detected.
- Flow Direc:
Flow direction output. Output is active in one flow direction and passive in the opposite.
- Empty Tube:
Empty tube signal is active, if no or less liquid is inside the meter tubes.