Rh eo n i k M esst ech n i k
RE V . L O ct o b er 2014
.
RHEONIK Coriolis Flowmeter
Operation Manual RHE 07, 08, 11
Page
25
d:
The d-values are given in promille (‰) of the time period of the base frequency.
Left is shown the phase shift of the last stored zeroing and right is the actual meas-
ured phase shift.
Press button
Vary
to see only the phase shift changes.
In the case of zero flow, this value
Vary
should be very low.
ADChannel1:
The Analog input value is displayed after conversion by the 10 bit analog/digital con-
verter. Value range is 0 - 1023 steps. This channel is in use for sensor temperature
measurement.
ADChannel2:
The second analog input channel has the same technical specifications as
ADChannel1
. It is a channel for special use (see refer. 2.1.3.5.2 analog input)
RTime:
Electronic run time counter in days and hours.
Mass:
Second, non resetable totalizer.
2.1.4.3. Sensor basic level programming (BASIS SETUP)
To enter this menu you have to input the password.
The password is:
- Press 3 times key #1
- Press 2 times key #2
- Press 1 time key
#3
After pressing a key an asterisk
*
is being displayed.
Inside this menu you have to enter sensor specific data like meter size, type of connec-
tion (serial or parallel) and maximum sensor operating temperature.