4. Warnings
The microprocessor also indicates warnings. The difference to errors is that warnings are
less dangerous than errors. For example there will be a warning when flow rate is above
100% sensor flow rate. But it is just a warning that tells you that meter accuracy could
be reduced in this range, yet hardware and software are working properly.
Possible warning codes are:
Code
Display
Description
_____________________________________________________________________
Warn1
Reset
Power failure possibly occurred, there was a processor reset.
Warn2
FlowRange
Flow rate is above maximum flow rate for this sensor size.
Reduce flow rate to have optimum accuracy.
Warn3
TempRange
Sensor temperature is more than allowable sensor
temperature (set in
MaxTemp
menu).
Reduce sensor temperature. The electrical installations inside the
sensor will be damaged.
Warn4
Drive
For a short time there was a lot of damping of the meter
oscillation (for instance, due to gas bubbles in liquids).
Warn5
OverflTot
There was a totalizer maximum count overflow, the
totalizer started again at zero.
5. Troubleshooting guide
5.1. Sensor voltages and resistances
There are four electrical circuits with which the sensor is connected to the transmitter
RHE. The sensor receives drive excitation from the transmitter and returns two AC Volt-
age signals back to the electronics. The fourth circuit is temperature measurement with a
PT100 temperature sensor.