Subsystems
CGS-240 Manual
6/15/2021
Page 67 of 148
P
RH
P
RL
Figure 6.25
Figure 6.26
T
R
SOL 8
injected into the evaporator, causing it to cool the circulating fluid. The pulse width of the valve directly controls
the amount of refrigerant injected, thereby affecting the temperature of the fluid. The valve is generally pulsed at
a fixed interval with varying pulse width. When the system requires cooling, the ECB’s DC CONTROL –
RLY 6
sources 24 VDC to the sat cool solenoid, causing it to open and inject refrigerant into the evaporator. An LED at RLY
6 also indicates the relay status. When the LED is ON, the valve is energized. Chamber fluid temperature is
measured by temperature probe, Tcl. Chamber temperature is measured by temperature probe, Tc.
6.4.5
EVAPORATORS
The refrigeration evaporators are brazed plate heat exchanger thermally coupling the sat and chamber fluids with
the R134A refrigerant system. One side of each evaporator is connected to the refrigeration system, while the
circulating fluid flows through the other side. As liquid refrigerant is injected into the evaporator, it expands into
gas. This expansion causes cooling (by absorbing heat), which lowers the temperature of the circulating fluid
flowing through the other side.
Saturator fluid flows through one evaporator to cool the saturator. Chamber fluid flows through the other
evaporator to cool the chamber.
6.4.6
REFRIGERATION PRESSURE SENSORS, PRL & PRH
The refrigeration low-side pressure sensor, P
RL
, measures the suction side
pressure of the refrigeration compressor. Readings from this sensor are in
gauge (relative) pressure. The sensor is powered from 12 VDC and has a 1-5
VDC output. The sensor is connected to the ECB at A/D INPUTS
–
AD1. Power
for the sensor is applied from AD1 12V+ and AD1
GND. The sensor’s voltage
output in the range of 1 to 5 VDC is connected to AD1 SIG.
The refrigeration high-side pressure sensor, P
RH
, measures the discharge side
pressure of the refrigeration compressor. Readings from this sensor are in
gauge (relative) pressure. The sensor is powered from 12 VDC and has a 1-5
VDC output. The sensor is connected to the ECB at A/D INPUTS
–
AD0. Power
for
the sensor is applied from AD0 12V+ and AD0 GND. The sensor’s voltage
output in the range of 1 to 5 VDC is connected to AD0 SIG.
6.4.7
EVAPORATOR BYPASS SOLENOID, SOL 8
The evaporator bypass solenoid, SOL 8, is used to cool the compressor during
times of high demand when cooling from high temperatures to low. It injects
refrigerant directly into the compressor suction line to overcome the heat
removed from the system in an attempt to keep the compressor cool. This valve
is actuated by a 24VDC pulse width modulated signal. When energized, the valve
opens allowing refrigerant to be injected directly into the suction line, cooling the
compressor. The pulse width of the valve directly controls the amount of
refrigerant injected, thereby affecting the temperature of the fluid. The valve is
generally pulsed at a fixed interval with varying pulse width. When the system
Содержание CGS-240
Страница 1: ...CGS 240 RH Systems LLC 1225 W Houston Ave Gilbert AZ 85233 Manual...
Страница 97: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 97 of 148 8 2 ILLUSTRATIONS Figure 8 1 CGS 240...
Страница 98: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 98 of 148 Figure 8 2...
Страница 99: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 99 of 148 Figure 8 3...
Страница 100: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 100 of 148 Figure 8 4...
Страница 101: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 101 of 148 Figure 8 5 Figure 8 6...
Страница 102: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 102 of 148 Figure 8 7 Figure 8 8...
Страница 103: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 103 of 148 Figure 8 9...
Страница 104: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 104 of 148 Figure 8 10...
Страница 105: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 105 of 148 Figure 8 11 Figure 8 12...
Страница 106: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 106 of 148 Figure 8 13 Figure 8 14...
Страница 107: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 107 of 148 Figure 8 15...
Страница 108: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 108 of 148 Figure 8 16...
Страница 109: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 109 of 148 Figure 8 17...
Страница 110: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 110 of 148 Figure 8 18...
Страница 111: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 111 of 148 Figure 8 19...
Страница 112: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 112 of 148 Figure 8 20...
Страница 113: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 113 of 148 Figure 8 21...
Страница 114: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 114 of 148 Figure 8 22...
Страница 115: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 115 of 148 Figure 8 23...
Страница 117: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 117 of 148 Figure 8 25...
Страница 118: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 118 of 148 Figure 8 26...
Страница 120: ...Drawings CGS 240 Manual 6 15 2021 Page 120 of 148...
Страница 121: ...Drawings CGS 240 Manual 6 15 2021 Page 121 of 148 Figure 9 3 Refrigeration System...
Страница 122: ...Drawings CGS 240 Manual 6 15 2021 Page 122 of 148 Figure 9 4 Pneumatic System...
Страница 123: ...Drawings CGS 240 Manual 6 15 2021 Page 123 of 148 Figure 9 5 Fluid System...
Страница 124: ...Drawings CGS 240 Manual 6 15 2021 Page 124 of 148 Figure 9 6 AC Distribution...
Страница 125: ...Drawings CGS 240 Manual 6 15 2021 Page 125 of 148 Figure 9 7 AD0 3 LVL4...
Страница 126: ...Drawings CGS 240 Manual 6 15 2021 Page 126 of 148 Figure 9 8 Power Distribution...
Страница 127: ...Drawings CGS 240 Manual 6 15 2021 Page 127 of 148 Figure 9 9 RLY 1 14...
Страница 128: ...Drawings CGS 240 Manual 6 15 2021 Page 128 of 148 Figure 9 10 PT 12 Temp Board...
Страница 129: ...Drawings CGS 240 Manual 6 15 2021 Page 129 of 148 Figure 9 11 UART 0 3...
Страница 130: ...Drawings CGS 240 Manual 6 15 2021 Page 130 of 148 Figure 9 12 Access to Presat Drain Fitting...
Страница 131: ...Drawings CGS 240 Manual 6 15 2021 Page 131 of 148 Figure 9 13 Access to Sat Drain Fittings...
Страница 132: ...Drawings CGS 240 Manual 6 15 2021 Page 132 of 148 Figure 9 14 Temperature Probes...
Страница 133: ...Drawings CGS 240 Manual 6 15 2021 Page 133 of 148 Figure 9 15 Pressure Transducers Manifold...