RAMP
® System Operation
Operator’s Manual, RAMP
®
Reader
34
From the SYSTEM Menu, press [▲] or [▼] until 9. SET LQC TIMER is displayed. Press [Enter]. The
following prompt is displayed, which is the system default:
To initiate the LQC TIMER/INTERVAL, press [▲] or [▼] until the desired LQC interval is displayed.
The range of the LQC Timer/Interval is 1-31 days.
Once the LQC interval is selected, press [Enter] to return to the SYSTEM Menu.
NOTE:
If [Options] is pressed instead of [Enter], the changes to the LQC interval will not
be accepted and the screen returns to the SYSTEM Menu.
To turn the LQC Timer off, press [▲] or [▼] until the message LQC TIMER IS OFF is displayed.
If an LQC Interval has been set, the Reader begins a countdown timer. When the timer reaches zero,
any subsequent sample result will be flagged with the message “WARNING LQC OVERDUE!” on
both viewing and printing sample results. If the LQC timer reaches zero and the Reader is displaying
the RUN/OPTIONS screen, the display will change to:
2.6.11 Reset LQC Timer
Resetting the LQC Timer starts a new countdown to the specified time interval.
From SYSTEM Menu, press [▲] or [▼] until 10. RESET LQC TIMER is displayed. Press [Enter] to
select this option. The Reader displays the following prompt along with a double beep:
Press [Enter] to select YES. The LQC Timer is reset to the specified interval and is confirmed by two
short beeps. The following prompt is displayed for 2 seconds, and the screen returns to the SYSTEM
Menu:
LQC TIMER HAS BEEN
RESET
RESET TIMER? > YES
NO
LQC NEEDED
PRESS KEY TO PROCEED
SET LQC TIMER
TO XX DAYS
SET LQC TIMER
(LQC TIMER IS OFF)
Содержание RAMP READER
Страница 11: ...Operator s Manual RAMP Reader Reader Figure 1 1 RAMP Reader Introduction 8...
Страница 12: ...Operator s Manual RAMP Reader Figure Reader Figure 1 2 RAMP Reader Rear Panel View Introduction 9...
Страница 14: ...Operator s Manual RAMP Reader RAMP System Operation Reader Figure 2 1 RAMP Reader Keypad System Operation 11...