System Specifications
The LabRAM I is equipped with the following external electrical connections:
•
AC Power Cord Inlet
•
Mini USB
•
Communication Connector for Automatic Vacuum System
All external connections are found in the back panel of the LabRAM I machine. See Figure 6 on
Page 17 for a detailed illustration.
Maximum Payload Capacity
1.1 lb (500 gram)
Sound Emitted
<70 dB at 3’ at 100
g
with an empty vessel
Vessel Temperature Monitoring
1 RTD monitored and recorded (optional)
Automatic/Programmable Vacuum
Yes (optional)
Data Logging
Yes
Computerized Operator Interface (HMI)
Yes
Weight
180 lb (82 kg) plus vessel holder fixture
Dimensions
18” wide x 26” deep x 25” high
Electrical @ 50/60 Hz, CE Certified
100-120 / 200-240 VAC
6.1 Voltage Level
The LabRAM I is designed to utilize single-phase 100-120 or 200-240 VAC 50Hz/60 Hz supply
for use worldwide.
6.2 Power Consumption
The LabRAM I is designed to be energy efficient by operating at a resonant condition when
mixing. Depending on the acceleration setpoint, vessel selection, and mix media, the LabRAM I
will draw between 80 and 1200 Watts
Table 6-1. LabRAM I System Specification.