English
21
Accessories
Parameter
Description
Range
Tube
Select the type of air tubing used by the patient. ClimateLineAir air tubing
is automatically detected when connected to the device.
SlimLine / Standard / 3m
AB filter
Select Yes if you attach an antibacterial filter.
No / Yes
Ext.
humidifier
Select Yes if you connect an external humidifier.
No / Yes
View
oximeter
Displayed at all times when an oximeter is connected.
18-300 BPM
0-100% SpO
2
Alarms
Alarm
Description
Range
High Leak*
Enable / disable the High Leak alarm.
Activates when a leak >40 L/min (0.7 L/sec) occurs for >10–30 seconds.
On / Off
Non-Vented
Mask
Enable / disable the Non-Vented Mask alarm.
Activates within 20–40 seconds when a non-vented mask is attached
during therapy.
Note: Use of supplemental oxygen with a vented mask may result in false
triggering of the Non-Vented Mask alarm.
On / Off
Low MV*
Sets the Low Minute Ventilation alarm.
Activates within 20–40 seconds after the measured level remains below
the set limit.
Note: Alarm may not trigger reliably when using a P10 mask.
Off / 1–10 L/min, 1 L/min
increments
Low SpO
2
Sets the Low SpO
2
alarm.
Activates when the SpO
2
value is below the set value for 20–40 seconds.
Only functional when an oximeter is connected.
When the Low SpO
2
alarm is set, the Oximeter sensor disconnected alarm
and Oximeter sensor failure alarm are also enabled.
Off / 70–95%, 1 % increments
Apnea
Sets the Apnoea alarm.
Activates when there is no inspiratory trigger (either patient or machine)
detected from the previous inspiration for the set interval. Two
consecutive inspirations (either patient or machine triggered) reset the
Apnoea alarm.
Off / 10–60 sec, 1 sec
increments
Alarm
Volume
Sets the alarm volume.
Low / Medium / High
*When enabled, SmartStart is automatically disabled.