English
27
4.
Partially block the open end of the air tubing with your hand.
The alarm activates within 10–20 seconds.
WARNING
In an environment where multiple devices are in use, the devices may have different alarm
settings.
Starting therapy
1.
Direct the patient to fit their mask.
2.
Direct the patient to press Start/Stop, or if the SmartStart feature is enabled, direct them to
breathe into their mask.
You will know that therapy is on when the
Monitoring
screen is displayed.
The pressure bar shows the inspiratory and expiratory pressures in green.
The green bar will expand and contract as the patient breathes in and out.
The screen will go black automatically after a short period of time. You can press Home or the dial to
turn it back on. If power is interrupted during therapy, the device will automatically restart therapy
when power is restored.
The AirCurve 10 device has a light sensor that adjusts the screen brightness based on the light in
the room.
Stopping therapy
1.
Direct the patient to remove the mask.
2.
Direct the patient to press Start/Stop, or if SmartStart is enabled, therapy will stop automatically
after a few seconds.
Notes:
•
If Confirm Stop is enabled, a message is displayed asking if you want to stop therapy. Turn the
dial to select
Yes
and then press the dial to stop therapy.
•
If Ramp Down is enabled, pressing Start/Stop will initiate Ramp Down.
The
Sleep Report
now provides a summary of the therapy session.