- 18 -
EN
Problem
Possible cause
Corrective action
Lamp does not
go on.
• Power cord not plugged in.
• Fluorescent tube defective.
• Faulty electronics.
• Fluorescent tube not prop
-
erly seated in the socket.
• Plug the power cord into the wall outlet.
• Replace the fluorescent tube (refer to Sec. 5.6).
• Call for service to replace the electronics, or refer to the spare
parts list.
• Refer to Sec. 5.6, “Fluorescent Lamp Installation”.
Compressed air
leaking from the
foot switch.
• Pneumatic hoses re
-
versed.
• Reverse the connections.
Foot switch fails
to switch on.
• No pressure at the com
-
pressed air connection.
• Pneumatic hoses are
kinked.
• Inspect the compressed air lines and/or the compressor.
• Reposition the hoses.
Foot switch hoses
cannot be discon-
nected from the
unit.
• Push-in connection is not
released (Fig. 1).
• Press the push-in connection ring on the hose side firmly
while pulling on the hoses.
We reserve the right to make technical change
Содержание Basic Master
Страница 3: ...1 2 3 4 5 6 7 8...
Страница 4: ...9 11 10 12 13 14 15 16...
Страница 5: ...17 18 19 20 21...
Страница 6: ......
Страница 42: ...42 RU 90002 8889 25 70 90002 8888 70 500 1 1 2 3 4 5 A apa ec epep He c pa a e a e a 5 6 5 6 1...
Страница 47: ...47 ZH 90002 8889 25 70 m 90002 8888 70 500 m 1 1 2 3 4 5 5 6 4 2 1...
Страница 48: ...48 ZH...
Страница 53: ...53 JA 15 2 4 6 90002 8889 25 70 m 90002 8888 70 500 m 1 mm 1 2 3 4 5...
Страница 54: ...54 JA 5 6 5 6 1...
Страница 59: ...59 KO 15 2 4 6 90002 8889 25 70 m 90002 8888 70 500 m 1mm 1 2 3 4 5 5 6 5 6...
Страница 60: ...60 KO Push in 1 Push in...
Страница 64: ...13b 10b 11b 14b 8 17 16 7 9a 9b 17 15 17 12 13a 11a 10a 14a...
Страница 67: ......