REBOTEC / BA No.: 02.09.136/2016-12
GB
- 6 -
Before setting up column (A), first unlock the wheel
brakes (F). To avoid damaging the paint coating on
the column, first straighten pedal (D).
Follow the instructions (X) on the bottom of the
column.
Next engage once again the castor locking brakes
(F)!
Step 6: Install the column
The main column (A) may now be set up. Insert
the M10 bolt from the side into the hole of the
bracket and the main column (A), all the way
through, and fasten it with the M10 cap nut.
Next from the front using an M12 Allen key, screw
on the grip handle firmly to the bracket. The grip
handle is now firmly connected and fixed to the
base member (C).
Cut off both cable ties after assembly. Remove the
protective film around the drive shaft and at the
end of the lifter arm.
Step 7: Install the lifter axis
To attach the drive motor
to the lifting axis, remove
the safety ring (1) and the
pin (2) from the borehole
of the lifting arm (B).
Next raise the lifter arm
(B) and align the borehole
of the lifting axis with the
lifter arm.
The lifter arm is fixed by
inserting the pin and in-
stalling the safety ring.
Be sure that the safety
ring is completely in-
stalled and closed
!
Step 8: Install the carrier
The carrier (G) (in the accessories) for the sling is
attached to the end of the lifter arm (B).
Loosen the pre-installed screw (1) at the end of the
lifter arm. Next insert the pre-assembled carrier
with the borehole facing upwards into the opening
of the lifter arm and tighten firmly.
Step 9: Install emergency OFF button
Before starting up the Pa-
tient Hoist, the emergency
OFF button must be deac-
tivated.
To do this, rotate the button
clockwise until it pops out.
If the rechargeable battery is connected to the
control unit, the charge status of the rechargeable
battery is displayed
(see page 7, section 10.3).
B
1
2
B
3
A
C
A
1
2
F
A
D
F
X
G
B
1